Dräger Evita 4 Instructions For Use Manual page 155

Intensive care ventilator. software 4.n
Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

Serial dead space Vds
Range
Resolution
Accuracy
Dead space ventilation Vds/V
Range
Resolution
Accuracy
Computed value displays
Compliance C
Range
Resolution
Range 0.7 to 99.9 mL/mbar
Range 100 to 200 mL/mbar
Accuracy
Resistance R
Range
Resolution
Range 3 to 99.9 mL/mbar
Range 100 to 200 mL/mbar
Accuracy
Leakage minute volume MV
Range
Resolution
Accuracy
T
0...90
Rapid Shallow Breathing RSB
Range
Resolution
Accuracy
Negative Inspiratory Force NIF
Range
Resolution
Accuracy
Curve displays
Airway pressure Paw (t)
Flow Flow (t)
Volume V (t)
Exp. CO
concentration FCO
2
*
C-values may be considerably falsified as spontaneous breathing increases;
compliance with the measuring accuracy therefore cannot be guaranteed for spontaneous breathing.
** R-values may be considerably falsified as spontaneous breathing increases;
compliance with the measuring accuracy therefore cannot be guaranteed for spontaneous breathing.
0 to 999 mL, BTPS
0.1 mL
±10 % of measured value or ±10 mL,
whichever is greater
T
0 to 99 %
1 %
±10 % of measured value
0.7 to 200 mL/mbar
0.1 mL/mbar
1 mL/mbar
±20 % of measured value*
3 to 200 mbar/L/s
0.1 mbar/L/s
1 mbar/L/s
±20 % of measured value**
leak
0 to 99 L/min, BTPS
0.1 L/min or for values less than 0.1 L/min: 0.01 L/min
± 18 % of measured value
approx. 35 s
0 to 9999 1 /(min x L)
1 /(min x L)
see measurement of V
–45 to 0 mbar
1 mbar
± 2 mbar
–10 to 100 mbar
–150 to 180 L/min
0 to 2000 mL
0 to 100 mmHg or
2
0 to 13 kPa
0 to 13 Vol.%
and f
T
or
Technical Data
155