3.7
After reprocessing
3.7.1
Assembling and fitting device-specific components
Prerequisites:
– All components are reprocessed and dry.
Assembling the expiratory valve or the neonatal expiratory valve
No. Expiratory valve
1
2
3
4
Procedure:
1. Make sure all components are completely dry; otherwise, this may impair proper
functioning.
2. Connect the flow sensor sleeve (1) to the expiratory valve or connect the muffler
(3) to the neonatal expiratory valve.
3. Attach the diaphragm (2) onto the edge of the expiratory valve housing.
4. Connect the water trap container (4).
Reprocessing instructions
1
4
Flow sensor sleeve
Diaphragm
-
Water trap container
|
Evita V series / Babylog VN series
2
3
Neonatal expiratory valve
-
Diaphragm
Muffler
Water trap container
Reprocessing
2
4
31