Preparing
Disinfecting/Cleaning
Basic device without ventilation hoses, gas
connection hoses and temperature sensor
Wipe disinfect
e.g. with Buraton 10 F or Terralin (Schülke & Mayr,
Norderstedt).
Comply with the manufacturer's instructions.
Cooling air filter, room air filter
Filters must be cleaned or replaced when soiled or at
the latest after 4 weeks, see page 144.
CO
Cuvette
2
Wipe off dirt, in particular inside and outside the
windows, with disposable tissue and cotton buds.
Disinfect with moist heat (93 °C/10 minutes) in a
cleaning and disinfecting machine. Use only cleaning
agent.
Or:
Disinfect in bath of disinfectant based on the listed
active substances, e.g. Cidex, Johnson & Johnson,
Norderstedt.
Or:
Steam-sterilise at 134 °C.
CO
Sensor
2
Wipe off dirt with cotton buds, in particular on the
windows of the CO
sensor.
2
Disinfect by wiping, e.g. with 70 % Ethanol.
140