Pressure Limited Ventilation (PLV)* – for manually
limiting pressure peaks using the Pmax pressure limit.
The tidal volume remains constant as long as the
pressure curve shows a plateau and the inspiratory flow
curve shows a brief flow pause between inspiration and
expiration.
To set the "Pmax" pressure limit, please refer to
page 115.
The value of Pmax is displayed as a dashed blue line in
the Paw (t) curve.
The "Inconstant volume" alarm is always active. It is
triggered automatically if the tidal volume V
longer be applied.
This visual and audible alarm can be suppressed with the
»Alarm Reset« key until the cause of the alarm is
remedied.
––––––––––––
* Please refer to page 160 for a detailed description of AutoFlow
and PLV.