Expiration Valve; Humidifier - Dräger Evita 4 Instructions For Use Manual

Intensive care ventilator. software 4.n
Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

Preparing
Dismantling

Expiration valve

1 Push the catch to the right while at the same time
2 pulling out the expiration valve.
If the expiration valve is fitted with an optional water trap:
Pull off the collecting jar.
Only strip down the expiration valve if badly soiled.
Unscrew the stopper by hand and remove together
with the diaphragm.
Do not disassemble the expiration valve any further.
Prepare the expiration valve for disinfecting and
cleaning by autoclave
and
prepare the expiration valve for hot steam sterilisation.
Place the open expiration valve in the basket so that it
cannot be damaged by other parts.

Humidifier

Dismantle in accordance with the specific Instructions
for Use and prepare for disinfecting/sterilising.
138
1
2