QUICK USER GUIDE FOR EVITA XL
Ventilating Infants and Small Children
Patient mode »Pediatrics«
Using infant size patient circuit connectors 6-42kg
1 Clamp hinged circuit support arm to rail on the left-hand
side and tighten screws.
• Connect ventilator circuit segments of appropriate lengths
(lengths suggested are in meters).
2 Install water trap(s) in vertical position.
Patient mode: »Adults« 42kg plus
1 Attach circuit support arm to the rail on the left-hand side of
the ventilator and tighten screws.
Connect ventilator circuit segments of appropriate lengths
(lengths suggested are in meters).
2 Install water trap(s) in vertical position if using circuit with them
3 Connect Y-piece, with the rubber sleeve of the Y-piece on
the inspiratory side (reusable silicone circuit).
Operating Concept
Control Panel
Its main elements are:
1 A large screen with all the information and controls needed for ventilation.
2 Fixed function keys ("hard" keys) major on-screen functions,
Alarm Silence« for silencing the
audible alarm for two minutes.
Alarm Limits« for adjusting
alarm limits.
Ventilator Settings« for setting ventilation mode and parameters.
Unassigned key for future functions.
» Sensor Parameter« for calibrating sensors.activating/deactivating monitoring
.
» System Setup« for configuringbsystem functions.
Start/Standby« for selecting standby mode or normal operation.
3 Central rotary dial knob for selecting, confirming settings on-screen.
The Screen...
.
1 Alarm messages
2 Operator prompts
Therapy status with ventilation mode,e.g. SIMV,
Patient mode, e.g. " Adult" for adults.
4 Waveforms, loops, and trends visualizing ventilation,
5 Presentation of set ventilation parameters
6 Essential, measured values of ventilation
7 Humidification type and status.
8 Touch-sensitive screen keys for the specific screen pages
9 Power supply indicator.
To select a screen page:
Touch the respective screen key:
Main« to select the main screen.
Values« to select a different group of measured
values in the field for numerical values.
»Data...« to display all measured values,
the logbook or trends on an additional screen
»Special Procedure... nebulization, O2 for suction
For full manual see
http://warney/ServiceManuals/D-H/DRAGER/

Advertisement

loading

Summary of Contents for Dräger EVITA XL

  • Page 1 QUICK USER GUIDE FOR EVITA XL Ventilating Infants and Small Children • Patient mode »Pediatrics« Using infant size patient circuit connectors 6-42kg 1 Clamp hinged circuit support arm to rail on the left-hand side and tighten screws. • Connect ventilator circuit segments of appropriate lengths (lengths suggested are in meters).
  • Page 2 To set the ventilation mode, ventilation parameters: 1 Press Ventilator Settings« or • touch one of the virtual screen knobs at the bottom of the screen. »Ventilator Settings« in the bottom part of the screen • Screen knobs and screen keys are displayed in the form of horizontal and vertical tabs as on a file catalog card.
  • Page 3 QUICK GUIDE TO DEVICE CHECK/LEAK TEST EVITA XL Checks of Readiness for Operation • To be performed before using on a patient, to confirm ventilator is working properly, ready for operation. • If the patient circuit is modified, the patient circuit leak test must be repeated.
  • Page 4 The following tests are performed during the device check: • System — Expiratory valve fit and throughput — Fit of flow sensor — Fit of neonatal flow sensor (for "NeoFlow" option) — Type of humifier — Completeness of patient hoses —...