Dräger Evita 4 Instructions For Use Manual page 120

Intensive care ventilator. software 4.n
Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

Configuration
Ventilation
Start-up values »V
, f« dependent on ideal weight.
T
The values are selected with reference to the Radford
nomogram:
Weight
Factory settings
kg
Tidal volume V
mL
3
20
15
110
65
450
100
700
The hospital-specific start-up values can be entered in
the table.
Start-up values »V
, f« dependent on patient mode.
T
Patient
Factory settings
mode
Tidal volume V
mL
Paed.
50
Adults
500
The hospital-specific start-up values can be entered in
the table.
To change the start-up values of »V
Touch the screen key of the parameter to be
changed.
Change value = turn rotary knob.
Confirm value = press rotary knob.
If you wish to return to the factory settings:
Touch »Dräger Reset« screen key.
120
Ventilation frequency f
T
bpm
30
26
13
10
Ventilation frequency f
T
bpm
29
12
, f«:
T
Hospital-specific settings
Tidal volume V
Ventilation frequency f
T
mL
......................................
......................................
......................................
......................................
......................................
......................................
......................................
......................................
Hospital-specific settings
Tidal volume V
Ventilation frequency f
T
mL
......................................
......................................
......................................
......................................
bpm
bpm