Preparation
Ventilation Adults and Children
Ventilation Adults and Children
From 100 mL tidal volume V
upwards
T
Patient mode: »Adults«
Do not use a heat and moisture exchanger at the
same time as a humidifier!
Risk of increased breathing resistance due to
condensation.
Connecting Aquapor humidifier
Prepare Aquapor following the relevant Instructions for
Use.
1 Hang Aquapor from rail by bracket and tighten
screws.
2 Insert elbow connector into Aquapor.
3 Insert the double connector into the elbow connector.
Fill Aquapor bowl to the upper mark with distilled
water.
Connecting ventilation hoses
Do not use antistatic or conductive hoses*.
Depending on the desired position of the ventilator in
relation to the bed, the hinged arm can be fitted to either
side of the machine.
Attachment on left-hand side:
4 Turn both ports to the left.
5 Turn Aquapor to the left.
The following description applies when the ventilation
hoses have been attached on the left-hand side.
______________
* DIN VDE 0750 Part 215:
The use of anti-static or electrically conductive material in the
breathing system of the lung ventilator is not considered conducive
to greater safety. On the contrary, the use of these materials
increases the danger of electric shock to the patient and of fire due
to the presence of oxygen.