Once the fault has been remedied, the alarm tone is
switched off. Caution and advisory messages disappear
automatically.
Warning messages (!!!) are displayed in the colour of the
status line and must be acknowledged:
1 Press »Alarm Reset« key.
The message is erased from the screen. However, it is
stored in Evita 4 and can be displayed with the logbook
function in the »Measured Values« screen page, see
page 85.

Cancel alarm tone

for max. 2 minutes:
g
2 Press the »
« key with the lit yellow indicator LED.
The acoustic alarm will be cancelled for 2 minutes. If
the fault that triggered the alarm is still not remedied,
the acoustic alarm starts up again. If you wish to
reactivate the acoustic alarm temporarily:
g
2 Press the »
« key (with the yellow LED now
switched off) again.
Acknowledge:
3 Press »Alarm Reset« to acknowledge the alarms that
can be suppressed with "Alarm Reset":
please refer to "Troubleshooting" on page 127.
J

Information

– For help in operating the system – with setting
instructions.
– For help in troubleshooting.
4 Press the »
« key: information is displayed in the
bottom line of the screen.
To erase the message:
4
Press the »
« key again.
D
D
D
Operation
In the Event of an Alarm
Evita 4
1
Evita 4
2
3
Evita 4
4
79

Advertisement