Dräger Evita 4 Instructions For Use Manual page 141

Intensive care ventilator. software 4.n
Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

Ventilation hoses, water traps and associated
jars, Y-piece, expiration valve (or, in the event
of severe fouling, their individual parts)
Disinfect with moist heat (93 °C/10 minutes) in
cleaning and disinfecting machine. Use only cleaning
agent.
After disinfecting with moist heat, steam-sterilise the
expiration valve and its individual parts at 134 °C.
Make sure that no liquid remains in the pressure
measuring line, since it might cause malfunction.
Or
If no washing machine is available:
Bath disinfecting e.g. with Sekusept (Henkel).
Comply with manufacturer's instructions.
Then rinse with clean water, preferably from a soft
water supply. Shake water out thoroughly, and leave
the products to dry.
Expiration valve and individual parts
after rinsing
steam-sterilise at 134 °C.
Or
Expiration valve
Rinse thoroughly with clear water, preferably from a
soft water supply.
Shake water out thoroughly.
After rinsing thoroughly, dry expiration valve.
After drying, steam-sterilise at 134 °C.
Ventilation hoses, water traps and
associated water jars, Y-piece, expiration
valve, temperature sensor
These parts can be steam-sterilised at 134 °C.
Preparing
Disinfecting/Cleaning
141