To set:
Touch the »Extra settings« screen key.
Touch the screen key corresponding to the desired
function.
For Flow Trigger and Sigh:
Touch the appropriate screen key.
Adjust the desired value = turn the rotary knob
Confirm the desired setting = press the rotary knob
Switch on the function = touch the »On« screen
button and press the rotary knob.
––––––––––––––––––––––––––––––––––––––––––––––––
IPPV can be configured for the additional ventilation
parameter Pmax.
Selecting »Pmax pressure limiting«, see page 115.
Pressure Limited Ventilation (PLV)* – for manually
limiting pressure peaks using the Pmax pressure limit.
The tidal volume remains constant as long as the pres-
sure curve shows a plateau and the inspiratory flow curve
shows a brief flow pause between inspiration and
expiration.
To set the "Pmax" pressure limit, please refer to
page 115.
The value of Pmax is displayed as a dashed blue line in
the Paw (t) curve.
The "Inconstant volume" alarm is always active. It is
triggered automatically if the tidal volume V
longer be applied.
This visual and audible alarm can be suppressed with
the »Alarm Reset« key until the cause of the alarm is
remedied.
––––––––––––
* Please refer to page 160 for a detailed description of AutoFlow
and PLV.
can no
T
®
Operation
Setting Ventilation Modes
55