10 Technical data
Output signal
Signal on alarm
Load
Switching output
Low flow cutoff
Galvanic isolation
96
10.1.4

Output

Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05...100 s),
full scale value selectable, temperature coefficient: typically 0.005% o.r./°C,
resolution: 0.5 µA.
< 700 Ω (for HART: R
• active: 0/4...20 mA, R
L
• passive: 4...20 mA, operating voltage 18...30 V DC, R
Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full-scale frequency 2...1000 Hz (f
on/off ratio 1:1, pulse width max. 10 s
• Pulse output: pulse value and pulse polarity selectable, max. pulse width adjustable
(0.05...2000 ms), above a frequency of 1 / (2x pulse width) the on/off ratio is 1:1.
• Current output → failsafe mode selectable (e.g. in accordance with NAMUR
Recommendation NE 43)
• Pulse/frequency output → failsafe mode selectable
• Status output → "not conductive" in the event of fault or power supply failure
Detailed data → Page 86
see "Output signal"
Status output:
Open collector, max. 30 V DC / 250 mA, galvanically isolated
Configurable for: error messages, flow direction, limit values
Switch points for low flow cutoff are selectable
All circuits for inputs, outputs, and power supply are galvanically isolated from each other.