9 Trouble-shooting
82
Type
Error message / No.
S
RANGE CUR.OUT n
!
# 351...354
S
RANGE FRQ.OUT n
!
# 355...358
S
RANGE PULSE n
!
# 359...362
S
SIGNAL TOO LOW
$
# 391
No. # 5xx
Application error
S
SW.-UPDATE ACT.
!
# 501
S
UP-/DOWNLOAD ACT
!
# 502
S
INITIAL. RUN
$
# 591
Cause
Current output:
Flow is out of range.
Frequency output:
Flow is out of range.
Pulse output:
Pulse output frequency is out of range.
Attenuation of acoustic measurement
section too high.
New amplifier or communication (I/O
module) software version is loaded.
Currently no other functions are pos-
sible.
Up- or downloading the device data
via configuration program. Currently
no other functions are possible.
Initialisation is running.
All outputs are set to 0.
Proline Prosonic Flow 90
Remedy / spare part
– Change the upper or lower range
values, as applicable.
– Increase or reduce flow, as applica-
ble.
– Change the upper or lower range
values, as applicable.
– Increase or reduce flow, as applica-
ble.
1. Increase the setting for pulse
weighting
2. When selecting the pulse width,
choose a value that can still be
processed by a connected counter
(e.g. mechanical counter, PLC
etc.).
Determine the pulse width:
– Variant 1: Enter the minimum
duration that a pulse must be
present at the connected coun-
ter to ensure its registration.
– Variant 2: Enter the maximum
(pulse) frequency as the half
"reciprocal value" that a pulse
must be present at the con-
nected counter to ensure its
registration.
Example:
The maximum input frequency of
the connected counter is 10 Hz.
The pulse width to be entered is:
1
--------------------- -
=
50 ms
2 10 Hz
3. Reduce flow.
– Check to see if the coupling fluid
must be renewed.
– It is possible that the fluid indicates
too much attenuation.
– It is possible that the pipe indicates
too much attenuation.
– Check the sensor spacing
(Installation dimensions).
– Reduce the number of traverses if
possible.
Wait until the procedure is finished.
The device will restart automatically.
Wait until the procedure is finished.
Wait until the procedure is complete.