10 Technical data
Reference operating
conditions
Maximum measured error
Repeatability
98
10.1.6

Performance characteristics

• Fluid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
For flow velocities of > 0.3 m/s and a Reynolds number of >10000, the system accuracy is:
• Pipe diameter DN < 50: ± 0.5 % o.r. ± 0.1 % o.f.s. *
• Pipe diameter 50 < DN < 200: ± 0.5 % o.r. ± 0.05 % o.f.s.
• Pipe diameter DN > 200: ± 0.5 % o.r. ± 0.02 % o.f.s.
o.r. = of reading
o.f.s. = of full scale value
* plastic pipes only
As standard, the system is dry-calibrated. The dry calibration procedure results in an additional
uncertainty of measurement. This measuring uncertainty is better than 1.5% typically. During dry
calibration, the characteristics of the pipe and the fluid are derived to calculate the calibration factor.
As verification of the accuracy, an accuracy report is offered as an option.
Accuracy is verified using a stainless steel pipe.
%
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.0
0
2
Fig. 53:
Max. measured error (wet calibration) in % of reading
a = pipe diameter DN > 200
b = pipe diameter 50 < DN < 200
c = pipe diameter DN < 50
max. ± 0.3% for flow velocities > 0.3 m/s
a
b
c
4
6
8
m/s
Proline Prosonic Flow 90
10
12
14
Endress+Hauser
A0001149