6 Commissioning
!
64
6.2

Commissioning via local display

6.2.1

"Sensor installation" Quick Setup menu

If the measuring device is equipped with a local operation, the sensor distance can be ascertained
using the "Sensor" Quick Setup menu (Fig. 44).
If a measuring device does not have a local operation, the individual parameters and functions must
be configured via the configuration program "ToF Tool - Fieldtool Package" (see Page 67).
ENDRESS+HAUSER
Esc
-
+
E
HOME-POSITION
Fig. 44:
"Quick Setup" menu for sensor installation
Note!
The display returns to the QUICK SETUP COMMISSIONING function cell if you press the ESC key
X
(
) during parameter interrogation.
Selection of the system units only influences the functions UNIT TEMPERATURE,
UNIT LENGTH and UNIT VELOCITY.
The POSITION SENSOR function only appears when the CLAMP ON option is set
in the MEASUREMENT function and the number of traverses is 2 or 4 in the
SENSOR CONFIGURATION function.
The WIRE LENGTH function only appears when the CLAMP ON option is set
in the MEASUREMENT function and the number of traverses is 1 or 3 in the
SENSOR CONFIGURATION function.
+ +
Quick Setup
E
E
+
Selection
Clamp On
Measurement
Sensor
Type
Sensor
Configuration
Pipe Standard
Nominal
Diameter
Pipe Material
Sound Velocity
Pipe
Pipe Diameter
Circumference
Wall Thickness
Liquid
Temperature
Sound Velocity
Liquid
Position
Sensor
Wire Length
Sensor
Distance
Proline Prosonic Flow 90
Setup
Sensor
Language
QS-Sensor
Units
Measurement
Insertion
Sensor
Type
Pipe Standard
Nominal
Diameter
Pipe Diameter
Circumference
Wall Thickness
Sensor
Distance
Path Length
F06-90xxxxxx-19-xx-xx-en-001
Endress+Hauser