6 Commissioning
!
64
6.2
Commissioning via local display
6.2.1
"Sensor installation" Quick Setup menu
If the measuring device is equipped with a local operation, the sensor distance can be ascertained
using the "Sensor" Quick Setup menu (Fig. 44).
If a measuring device does not have a local operation, the individual parameters and functions must
be configured via the configuration program "ToF Tool - Fieldtool Package" (see Page 67).
ENDRESS+HAUSER
Esc
-
+
E
HOME-POSITION
Fig. 44:
"Quick Setup" menu for sensor installation
Note!
The display returns to the QUICK SETUP COMMISSIONING function cell if you press the ESC key
X
(
) during parameter interrogation.
➀
Selection of the system units only influences the functions UNIT TEMPERATURE,
UNIT LENGTH and UNIT VELOCITY.
➁
The POSITION SENSOR function only appears when the CLAMP ON option is set
in the MEASUREMENT function and the number of traverses is 2 or 4 in the
SENSOR CONFIGURATION function.
➂
The WIRE LENGTH function only appears when the CLAMP ON option is set
in the MEASUREMENT function and the number of traverses is 1 or 3 in the
SENSOR CONFIGURATION function.
+ +
Quick Setup
E
E
+
➀
Selection
Clamp On
Measurement
Sensor
Type
Sensor
Configuration
Pipe Standard
Nominal
Diameter
Pipe Material
Sound Velocity
Pipe
Pipe Diameter
Circumference
Wall Thickness
Liquid
Temperature
Sound Velocity
Liquid
Position
➁
Sensor
➂
Wire Length
Sensor
Distance
Proline Prosonic Flow 90
Setup
Sensor
Language
QS-Sensor
Units
Measurement
Insertion
Sensor
Type
Pipe Standard
Nominal
Diameter
Pipe Diameter
Circumference
Wall Thickness
Sensor
Distance
Path Length
F06-90xxxxxx-19-xx-xx-en-001
Endress+Hauser