4.2.3

HART connection

Users have the following connection options at their disposal:
• Direct connection to transmitter by means of terminals 26 / 27
• Connection by means of the 4...20 mA circuit.
Note!
• The measuring loop's minimum load must be at least 250 Ω.
• After commissioning, make the following settings:
– CURRENT SPAN function → "4...20 mA HART" or "4...20 mA (25 mA) HART"
Connecting the HART handheld communicator
For connecting, see also the documentation issued by the HART Communication Foundation, and
in particular HCF LIT 20: "HART, a technical summary".
4
3
Fig. 36:
Electrical connection of the HART handheld communicator:
1 = HART communicator, 2 = power supply, 3 = shield, 4 = other evaluation devices or PLC with passive input
Connecting a PC with operating software
A HART modem (e.g. Commubox FXA 191) is required for connection of a personal computer
running operating software (e.g. "ToF Tool - Fieldtool Package").
See also the documentation issued by the HART Communication Foundation, and in particular
HCF LIT 20: "HART, a technical summary"
.
1
RS 232
4
3
Fig. 37:
Electrical connection of a PC with operating software
1 = PC with operating software, 2 = power supply, 3 = shield, 4 = other switching units or PLC with passive input,
5 = HART modem, e.g. Commubox FXA 191
6
9
Page
Up
Bksp
Delete
Page
On
# % &
A B C
D E F
1
2
3
Copy
Paste
Hot Key
G H I
J K L
M N O
4
5
6
Insert
+ Hot Key
P Q R S
T U V
W X Y Z
7
8
9
, ( ) '
_ < >
+ * /
.
0
-
375
FIELD COMMUNICATOR
1
³
W
250
5
³
W
250
4 Wiring
-27
2
+26
-27
2
+26
A0001136
A0001137
39

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents