Proline Prosonic Flow 90
Installation instructions
Inlet and outlet runs
Connecting cable length
Ambient temperature
Storage temperature
Degree of protection
Endress+Hauser
10.1.7

Operating conditions

Installation
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions → Page 14 ff.
Clamp-on version → Page 15
Insertion version → Page 16
Shielded cables are offered in the following lengths:
5 m, 10 m, 15 m and 30 m
Route the cable well clear of electrical machines and switching elements.
Environment
• Transmitter Prosonic Flow 90:
–20...+60 °C
optionally: –40...+60°C
!
Note!
At ambient temperatures below –20°C the readability of the display may be impaired.
• Flowrate measuring sensors Prosonic Flow P (clamp-on):
–40...+80 °C / 0...+170 °C
• Flowrate measuring sensors Prosonic Flow W (clamp-on):
–20...+80 °C
• Flowrate measuring sensors Prosonic Flow U (clamp-on):
–20...+60 °C
• Flowrate measuring sensors Prosonic Flow W (Insertion):
–40...+80 °C
• Sensor cable PTFE: –40...+170 °C; Sensor cable PVC: –20...+70 °C
• In heated piping or piping conveying cold fluids, it is always permissible to insulate the piping
completely with the mounted ultrasonic sensors.
• Install the transmitter at a shady location. Avoid direct sunlight, particularly in warm climatic
regions.
The storage temperature corresponds to the operating temperature range of the measuring transmit-
ter and the appropriate measuring sensors and the corresponding sensor cable (see above).
• Transmitter Prosonic Flow 90:
IP 67 (NEMA 4X)
• Flowrate measuring sensors Prosonic Flow P (clamp-on):
IP 68 (NEMA 6P)
• Flowrate measuring sensors Prosonic Flow W (clamp-on):
IP 67 (NEMA 4X)
• Flowrate measuring sensors Prosonic Flow U (clamp-on):
IP 54
• Flowrate measuring sensors Prosonic Flow W (Insertion):
IP 68 (NEMA 6P)
10 Technical data
99