Trouble-shooting

9.1

Trouble-shooting instructions

Always start trouble-shooting with the checklist below, if faults occur after start-up or during oper-
ation. This takes you directly (via various queries) to the cause of the problem and the appropriate
remedial measures.
Check the display
No display visible and no output
signals present.
No display visible, but output
signals are present.
Display texts are in a foreign
language.
Measured value indicated, but
no signal at the current or pulse
output
Error messages on display
Errors that occur during commissioning or measuring are displayed immediately.
Error messages consist of a variety of icons. The meanings of these icons are as follows:
– Error type: S = system error, P = process error
$
– Error message type:
= fault message, ! = notice message
– S.VELOC RANGE = error designation (e.g. sound velocity outside measuring range)
– 03:00:05 = duration of error occurrence (in hours / minutes /seconds)
– # 491 = error number
"
Caution!
• See the information on Page 47 ff.!
• The measuring system interprets simulations and measured-value suppression as system errors, but displays them as
information messages only.
Error number:
No. 001
399
No. 501
799
Error number:
No. 401
499
Other error (without error message)
Some other error has occurred.
1. Check supply voltage
Terminal 1, 2
2. Check device fuse
Page 92
85...260 V AC: 0.8 A slow-blow / 250 V
20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
3. Electronics defective
Order spare part
1. Check whether the ribbon-cable connector of the display module is correctly
plugged into the amplifier board
2. Display module defective
3. Electronics defective
Order spare part
Switch off power supply. Press and hold down both the OS keys and switch on the
measuring device. The display text will appear in English (default) and is displayed
at maximum contrast.
Electronics PCB defective
Order spare part
System error (device error) has occurred → Page 80
Process error (application error) has occurred → Page 84
Diagnosis and remedial measures