LASer:LIV:OUTPUT <nrf value>
Action
Parameters
Notes
Examples
LASer:LIV:OUTPUT?
Action
Parameters
Notes
Examples
03_10
Enables or disables fast LIV current source output.
An <nrf value>; 1 = on; 0 = off.
There is a short delay between the LIV output is enabled and the shorting relay is
opened. Current quickly ramps up from the start current to the stop current. The basic
ramping time (without setting the delay time) for each step is 10 ms for CW. The basic
ramping time for each step is dependent on the pulse width and duty cycle in QCW. The
fast LIV current source output will automatically be turned off when the stop current is
reached. The customer can also choose to turn off the fast LIV in the middle of the LIV
running.
"LAS:LIV:SET 0, 7, 0.01, 0.001; LAS:LIV:OUTPUT ON" - action: the start current is set to
0A, the stop current is set to 7A, the step current is set to 0.01A, the delay time in each
step is 1 ms, and the output is enabled.
"las:liv:output 0" - action: LIV output is disabled.
Returns status of LIV output.
None.
None.
"LAS:LIV:OUT?" - response: 0 meaning the LIV output has been disabled.