ILX Lightwave LDX-36000 Series User Manual page 115

High power laser diode current source
Table of Contents

Advertisement

LASer:ENABle:OUTOFF?
Action
Parameters
Notes
Examples
03_10
Returns the value of the Output Off Enable Register.
None.
The response is the sum of the enabled bits and must be a value between 0 and 65535
as detailed below.
1
Laser Current Limit
*2
Laser Voltage Limit
4
High Temperature Limit
*8
Laser Open Circuit
*16
Interlock #1 Open
*32
Interlock #2 Open
64
Temp Sensor Open
128
Temp Sensor Shorted
The enabled or disabled events can be set via the LASer:ENABle:OUTOFF command.
The output status can be monitored by using the LASer:EVEnt? query.
The factory default value for this register is #B1111110000111110, or #HFC3E, or 64574
decimal.
Bits marked by asterisks (*) are permanently enabled and cannot be changed.
Bits marked with a dagger (
"LAS:ENAB:OUTOFF?" -response: "1086" means, that either the laser voltage limit,
high temperature limit, laser open circuit, the opening of interlocks #1 or #2, or the
hardware current limit will force the laser output off.
C O M M A N D R E F E R E N C E
Command Reference
256
Output On
512
Bad Steinhart-Hart Conversion
*1024
Hardware Current Limit
2048
* AC Power Failure
*4096
Laser Open Circuit 2
*8192
Power Supply Failure
*16384
Power Supply Voltage Limit
*32768
Pass Element Power Limit
) are permanently disabled and cannot be changed.
5
C H A P T E R
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT
99
LDX-36000 Series

Hide quick links:

Advertisement

Table of Contents
loading