ILX Lightwave LDX-36000 Series User Manual page 124

High power laser diode current source
Table of Contents

Advertisement

5
C H A P T E R
LASer:MODE
Action
Parameters
Notes
Examples
LASer:MODE:CW
Action
Parameters
Notes
Examples
LASer:MODE:HPULSE
Action
Parameters
Notes
Examples
108
LDX-36000 Series
C O M M A N D R E F E R E N C E
Command Reference
Returns the selected laser output mode.
None.
The response is the same as the labels of the illuminated mode LEDs.
"LAS:MODE?" - response: "CW", means the output mode is CW. The CW LED is
illuminated in the mode box on the front panel along with the CW LED in the CW Mode
box.
"Las:Mode?" - response: "TRIG", means the output mode is QCW-Trig. the QCW LED is
illuminated in the Mode box on the front panel along with the Trig LED in the QCW Mode
box.
Places the LDX-36000 in CW output mode.
None.
If the output was enabled when the command is given, the output will be disabled prior to
changing modes. If the command is given when the instrument is already in the
specified mode, output will be disabled.
"LAS:MODE:CW" -action: output is disabled if it was on and the output mode is changed
to CW.
Places the LDX-36000 in hard pulse mode
None.
If the output was enabled when the command is given, the output will be disabled prior to
changing modes. If the command is given when the instrument is already in the
specified mode, the output will be disabled.
In this mode, pulse width, frequency, and duty cycle are controlled either from the front
panel or via GPIB.
"LAS:MODE:HPULSE" - action: output is diabled if it was on and the output mode is
changed to hard pulse mode when the instrument is in QCW mode.
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT