ILX Lightwave LDX-36000 Series User Manual page 136

High power laser diode current source
Table of Contents

Advertisement

5
C H A P T E R
Temperature?
Action
Parameters
Notes
Examples
TERM <nrf value>
Action
Parameters
Notes
Examples
TERM?
Action
Parameters
Notes
Examples
120
LDX-36000 Series
C O M M A N D R E F E R E N C E
Command Reference
Returns the measured temperature in °C.
None.
Temperature measurement must be enabled by setting all three Steinhart-Hart thermistor
constants to nonzero values before this query may be used. If temperature is queried
while measurement is disabled, the last valid temperature measurement will be returned.
"T?" -response: "27.5" means the measured laser temperature is 27.5°C.
Specifies the terminator to be used in GPIB communications.
One Boolean <nrf value> specifying whether or not a carriage return <CR> will be added
to the GPIB terminator. 0 = FALSE, non-zero = TRUE.
An altered terminator will be in the form <CR><NL><^END>. This technically takes the
LDX-36000 out of the IEEE-488.2 specification, but may be done for convenience when
using non-standard GPIB controllers. This termination will be sent with all output until
the TERM 0 command is sent, or the instrument is turned off.
"Term 1" -action: temporarily sets <CR><NL><^END> as the output terminator.
"Term 0" -action: returns output terminator to <NL><^END>.
Returns whether or not a carriage return has been added to the GPIB terminator.
None.
An altered terminator will be in the form <CR><NL><^END>. This termination will be
sent with all output until the TERM 0 command is sent, or the instrument is turned off.
"Term?" -response: "1", the output terminator has been temporarily set to
<CR><NL><^END>.
"Term?" -response: "0", the output terminator is the IEEE-488.2 standard <NL><^END>.
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT
C
OMMON
F
P
RONT
ANEL
D
D
EVICE
EPENDENT