Error Messages; Table 6.2 Error Code Classifications - ILX Lightwave LDX-36000 Series User Manual

High power laser diode current source
Table of Contents

Advertisement

Error Messages

In the event of a hardware error condition, error messages are displayed on the
Left-hand display or Display 1. In most cases, the error message appears for
three seconds. In the case of multiple error messages, the instrument will
sequentially show each message for three seconds. In addition to the hardware
errors, GPIB errors may be read via the ERR? query.
Table 6.2 lists the numerical error ranges by function. Table 6.3 contains all of the
potential error messages. Not all of these messages may appear on the front
panel display; some refer to GPIB activities only.
In remote operation, the errors may be read by issuing the ERR? query. When
this is done, all of the error messages which are resident in the error queue are
returned (up to 10 may be stored). Reading the error queue via GPIB clears the
error queue.

Table 6.2 Error Code Classifications

Error Code Range
E-001 to E-099
E-100 to E-199
E-200 to E-299
E-300 to E-399
E-400 to E-499
E-500 to E-599
03_10
C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Note: Error codes not listed are reserved for future design use.
Error Messages
Area of Operation
Internal Program Errors
Parser Errors
Execution Control Errors
GPIB Errors
Not Applicable
Output Control Errors
6
C H A P T E R
135
LDX-36000 Series