LASer:CAL:MDI <nrf value>,<nrf value>
Action
Parameters
Notes
Examples
LASer:CAL:MDI?
Action
Parameters
Notes
Examples
LASer:CAL:THERMI <nrf value>,<nrf
value>
Action
Parameters
Notes
Examples
78
LDX-36000 Series
C O M M A N D R E F E R E N C E
Command Reference
Enters the slope and offset calibration constants for photodiode current measurement.
Two <nrf values>. The first parameter is the setpoint slope and the second is the
setpoint offset.
This command allows for calibration and linearization of the photodiode current
measurement. The complete calibration procedure is outlined in Chapter 6.
Default values for slope and offset are 1 and 0, respectively.
"las:cal:mdi 1.004874, 0.049776" -action: Adjusts the internal A/D calibration to linearize
the voltage limit by entering a new slope and offset of 1.004874 and 0.049776,
respectively.
"Las:Cal:MDI 1.0, 0.0" -action: Resets the photocurrent measurement calibration
constants to default values.
Returns the current photodiode current measurement calibration constants stored within
the instrument.
None.
Two values are returned by this query, a slope and offset value, respectively. This
command is used to verify successful calibration by reading back the stored constants.
This procedure is outlined in Chapter 6.
"las:cal:mdi?" -response: 1.0, 0.0, meaning a slope of 1.0 and an offset of 0.0 had been
previously entered as part of the calibration process
Enters the slope and offset calibration constants for thermistor current.
Two <nrf values>. The first parameter is the slope and the second is the offset.
This command allows for calibration of the thermistor current output. The complete
calibration procedure is outlined in Chapter 6.
Default values for slope and offset are 1 and 0, respectively.
"las:cal:thermi 1.004874, 0.049776" -action: Adjusts the internal D/A calibration to adjust
the thermistor current by entering a new slope and offset of 1.004874 and 0.049776,
respectively.
"Las:Cal:ThermI 1.0, 0.0" -action: Resets the photocurrent measurement calibration
constants to default values.