Appendix

Non-condensing Conditions

Non-condensing Conditions
The Agilent 3100 OFFGEL Fractionator is specified to be operated only under
non-condensing conditions and with a cooling platform temperature of
maximum 10 °C below ambient temperature.
For example, assuming a typically experimental setup of 20 °C on the cooling
platform and an ambient temperature of 25 °C, the humidity must be below
73 %. A plot of humidity versus ambient temperature for the cooling platform
at 20 °C, 25 °C and 30 °C is depicted in
conditions the humidity has to be below the plotted curves.
Figure 1
154
Plot of humidity versus ambient temperature assuming the cooling platform to
be set at 20 °C, 25 °C or 30 °C. For non-condensing condition the humidity has
to be below the plotted curves.
Figure
1. For non-condensing