the port at the rear of instrument. Upon recognition the browser interface will
show the USB stick symbol in the status screen. In case the USB-memory stick
is FAT-32 formatted the instrument will attempt to format it to FAT-16. The
file browser of your computer operating system also allows you to format USB
sticks. FAT-16 is typically noted simply as FAT.
Steps for Firmware Change via USB
1 Download the Agilent 3100 OFFGEL Fractionator firmware from the
2 Double-click the exe-file to extract the content into any folder on your hard
3 Create a folder on the USB stick with exactly this path: X:\OFFGEL\
4 Move the extracted firmware file to the new folder on the USB stick. Only
5 Power the 3100 OFFGEL Fractionator off.
6 Insert the prepared USB stick to the USB port of the instrument.
7 Power the Agilent 3100 OFFGEL Fractionator on, during the restart the
8 Allow the instrument to perform the Firmware update. A typical time for
Do not disturb the firmware process. Any interruption will lead to incomplete firmware
N O T E
programming. Instruments in such status are not functional.
OFFGEL User Manual
Agilent web site
(http://www.chem.agilent.com/scripts/DownloadsUtilities.asp) onto your
PC.
disc drive. The Main Firmware file has a name starting with the instrument
number, revision description and the suffix .dlb. An example for a typical
Main Firmware file name is: 3100A_B630_002.dlb.
UPDATE. Please use capital letters for the path. X: represents the drive
letter of your USB stick and is a variable.
exactly this file should be present in the folder.
instrument will check the USB stick folders for relevant files and perform
an update in case the installed firmware has a lower revision.
this is 10 minutes. During this time the red status LED will blink.
After a successful update (or downgrade) of main or resident firmware, the
3100 OFFGEL fractionator restarts automatically. The instrument restart is
indicated by a toggling of the three LED colors. In case main and resident
firmware are changed in one attempt wait until the second cycle of red
blinking LED and a toggling of the three LED colors is complete.
Maintenance
6
Firmware Update
89