Status Bars; Error Messages - Agilent Technologies 3100 OFFGEL Fractionator User Manual

Table of Contents

Advertisement

Firmware download Tool and update to the latest firmware revision. The
resident section of the firmware is identical in many Agilent devices and
modules. Its properties include:
• The complete communication capabilities (CAN, LAN and RS-232C)
• Memory management
• Ability to update the firmware of the "main system"

Status Bars

The Agilent 3100 OFFGEL interface displays several status bars (see
Functions"
Table 9
Status bar title
Cover open
Temp
Wait Temp I or
Wait Temp II

Error Messages

Table 10
Error message
USB drive full. Exchange USB drive or delete
files.
File export failed. The selected files cannot get
copied to USB drive. Check if drive is full.
OFFGEL User Manual
on page 42).
Status bar title overview - LED is yellow
Description
The top cover of the instrument is
open.
The set temperature of the cooling
platform has not been reached.
The set temperature of the cooling
platform has not been reached;
however, a fractionation was already
started on one of the trays.
Error messages due to wrong operation
Troubleshooting the Instrument
Comment
A running fractionation will be
paused if the top cover is opened
You can start a fractionation;
however, the status bar will switch
to Wait Temp I/II.
Once the cooling platform
temperature is reached the
fractionation will be started
automatically.
Solution
Exchange the USB drive or delete files on the
USB drive.
Exchange the USB drive or delete files on the
USB drive.
Troubleshooting
7
"General
93

Advertisement

Table of Contents
loading