Error message
Cover sensor failed (EE 20300).
Temperature sensor failed (EE
20301).
Fan failed (EE 00068).
Leak (EE 00080, EE 00081, EE
00082 or EE 00083).
High voltage (EE 20305).
Peltier Error (EE 20304).
In case you wish to call Agilent Service for support on the 3100 OFFGEL fractionator provide
N O T E
a detailed error description, exported results from your latest sample runs, diagnosis
buffers, content of the instruments log book and firmware revision.
OFFGEL User Manual
Error messages hardware related - LED is red (continued)
Comment
Opening and closing the top
cover is no longer recognized.
The voltage did not reach the
set point.
The voltage or currents of the
Peltier elements are out of
limits.
Troubleshooting
Troubleshooting the Instrument
Action
• Check that the top cover
closes tightly.
• The cover sensor might be
defective. Call Agilent
Technologies.
• The temperature sensor is
defective. Call Agilent
Technologies.
• The fan might be defective.
Call Agilent Technologies.
• The leak sensor is defective.
Call Agilent Technologies.
• Check for condensation or
dirt on anode connector,
electrodes or tray. Clean the
the parts as described in
"Cleaning the Agilent 3100
OFFGEL Fractionator"
page 70.
• The high voltage module
might be defective. Call
Agilent Technologies.
• A Peltier element might be
defective. Call Agilent
Technologies.