Be sure to not to confuse the LAN connector with the CAN connector. The web browser
N O T E
access to the instrument will not work when connecting it to the CAN connector.
3 Connect the power cable to the instrument and to the line power.
4 Verify configuration dip-switches
5 Switch on the instrument.
Figure 8
The instrument LED is usually yellow, which indicates that the cooling
platform is cooling down to the specified temperature (typically 20°C). Once
the temperature has been reached, the LED is switched off.
To test the instrument for proper functionality, it is recommended to perform the tests as
N O T E
specified in chapter
OFFGEL User Manual
Switches N° 1 to 6 must be down
Switches N° 7 and 8 must be up
Configuration switches of the Agilent 3100 OFFGEL Fractionator
"Troubleshooting the Instrument"
Setting up the Agilent 3100 OFFGEL Fractionator
(Figure
8):
on page 92.
Installation