If a worksheet is manually deleted or the data file is reloaded, the corresponding graph is
N O T E
deleted. If a plot to a certain worksheet is repeatedly generated, the previous plot version is
deleted.
Table 5
Table 5
Run Parameter
VoltTT, PowerTT
kVHours, Volt, Power
Temp
Top Cover
CurrA, CurrB, CurrC, ...
µAhA, µAhB, µAhC, ...
OFFGEL User Manual
gives details of the different run parameters:
Run parameters
Reviewing Result Files
Description
Fixed settings of Volt and Power of the method
that was used.
kVHours, Volt, Power measured during the
fractionation for the complete tray.
Temperature of the cooling platform during the
fractionation.
The events or time intervals when the top cover
was opened during a fractionation are marked
with the red diamond symbol on the X-axis. If
there are no Top Cover events registered, the Top
Cover check box is automatically unchecked.
Currents measured for individual tray lanes
(A-H).
Measured µA-hours acquired for individual tray
lanes (A-H).