Agilent Technologies 3100 OFFGEL Fractionator User Manual page 44

Table of Contents

Advertisement

5
Operation
Status tab
fractionation. Not ready indication on for example tray temperature will lead
to a waiting step. An open lid will cause the not ready information and
application of the voltage is done only if the lid is finally closed, see also
Table 8
indicated by a blue tray corona around the respective tray-shaped interface.
An Agilent 3100 OFFGEL Fractionators method generally consist out of two
method steps which are the focusing and the hold phase. Methods are
described in more detail under
parameter shown in the status screen are the specified limits for volt hours,
time, voltage, power and current of the loaded method. The bar graph on the
left side visualizes the actual values online. The bar size is in relation to the
set point; in addition, the numeric value is displayed by a tool tip if the mouse
pointer is place over the respective bar. The bar graph labeled with letters
from a to h on the right side visualizes the current measured as actual per
lane for each strip online. Click into the lane area to swap from displaying
Current in µA to Charge in mAh.
Modification of a currently loaded method can be done by clicking onto the
Edit button or by clicking into the respective parameter area of a tray. A
dialog will appear and allows changing the parameter of the current method
step, to directly apply it to the ongoing fractionation or to cancel this attempt (
see
method name and denotes that the method was modified during the run.
Figure 12
44
on page 92. The active Run status for each tray is additionally
Figure 12
on page 44). Upon modification an asterisk (*) occurs next to the
Edit a current method
"Working with Methods"
on page 46. The run