Troubleshooting the LAN Connection
If your web interface to the Agilent 3100 OFFGEL Fractionator does not work
correctly check for disconnected LAN cables at the rear of the instrument.
Make sure you are using the LAN connector and not the CAN connector.
N O T E
If the problem still remains, then:
• Power down the Agilent 3100 OFFGEL Fractionator and wait 1 minute and
• Check the positions at the rear of the instruments according to the
If the problem still remains and you were had the instrument in a local
network, then:
• Set up the instrument in a direct LAN connection to a PC. Install a cross
Call Agilent Technologies if the problem cannot be solved.
OFFGEL User Manual
then restart.
installation instruction in chapter 3.
over LAN cable and change the configuration dip-switches 7 and 8 in the
rear of the instrument to On. Follow the instruction in
connection is reestablished review and correct the IP address in the Config
tab of the web interface.
Troubleshooting
Troubleshooting the LAN Connection
Chapter
7
4. Once the
101