2
Site requirements and specifications
Site Requirements for the Agilent 3100 OFFGEL Fractionator
Site Requirements for the Agilent 3100 OFFGEL Fractionator
Power Considerations
The Agilent 3100 OFFGEL Fractionator power supply has wide ranging
capabilities and accepts any line voltage in the range 100–240 V with a line
frequency of 50–60 Hz (see also on page 25 for more details). Consequently,
there is no voltage selector in the rear of the instrument.
Shock hazard or damage to your Agilent 3100 OFFGEL Fractionator may result if the
C A U T I O N
Agilent 3100 OFFGEL Fractionator is connected to a line voltage higher than specified.
Bench Space
The Agilent 3100 OFFGEL Fractionator requires sufficient desk or laboratory
bench space; dimensions and weights are shown in
Approximately 15 cm (6 inches) of clear space at the rear of the instrument for
electric connections is needed.
Environment
The Agilent 3100 OFFGEL Fractionator will work within the specifications of
ambient temperatures and relative humidity described in
Do not store, ship or use the Agilent 3100 OFFGEL Fractionator under conditions where
C A U T I O N
temperature fluctuations could cause condensation within the Agilent 3100 OFFGEL
Fractionator. Condensation will damage the electronics. If the Agilent 3100 OFFGEL
Fractionator was shipped in cold weather, leave it in its box and allow it to warm up
slowly to room temperature to avoid condensation.
22
Table
on page 25.
Table 2
on page 25.
OFFGEL User Manual