Lan; Tested Configurations - Agilent Technologies 3100 OFFGEL Fractionator User Manual

Table of Contents

Advertisement

LAN

The 3100 OFFGEL Fractionator comes with a default IP address
(192.168.254.11) that cannot be changed. The instrument can also be assigned
a user settable, static IP address. This custom IP address can be set after an
initial connection via the default address. Configuration dip-switches at the
instrument rear allow to select which of the two addresses are used by the
3100 OFFGEL Fractionator. This configuration is applied when the instrument
is powered on the next time.
A detailed description on the steps for instrument setup is given under
Chapter
The instrument is not enabled for DHCP (Dynamic Host Configuration Protocol). Therefore
N O T E
an automated assignment of network configurations from a server is not possible. For
incorporation into a local network a static IP address is needed.

Tested Configurations

The configuration of the typical bundle equipment (Part Number G3100AA) of
the system is described below. It was successfully tested and consists of
• a 3100 OFFGEL Fractionator
• a Netbook HP Mini 5102, Operating system Windows 7 Starter Edition,
• a connection by a Cross-Over LAN cable
Another successfully tested configuration consisted of:
• a 3100 OFFGEL Fractionator with wired LAN connection to a local Network
• Apple iPod Touch 8G (FW Revision 3.1.2, including a Safari Browser) with
OFFGEL User Manual
4, "Installation," starting on page 31.
Internet Explorer 8.0
wireless LAN access to the local Network.
PC, Browser and LAN Requirements
PC, Browser and LAN Requirements
3