Troubleshooting The Instrument; Status Indicators; Status Led - Agilent Technologies 3100 OFFGEL Fractionator User Manual

Table of Contents

Advertisement

7
Troubleshooting

Troubleshooting the Instrument

Troubleshooting the Instrument

Status Indicators

Status LED

The instrument has a status LED on the front top right position (see
on page 17). With the power switched on, the LED can have the following
states:
Table 8
LED Status
Green
Yellow
Red
Off
While the main firmware is downloaded to the instrument, the status indicator LED on the
N O T E
front of the instrument front will blink in red.
The status LED will blink red in case the instrument uses the resident
firmware only. This might happen if the main firmware installed to the
instrument encountered an error or is corrupt. In such case only limited
functionality is available. Please restart the instrument. If the blinking LED is
still present after an instrument restart, connect to the instrument with the
92
LED status overview
Description
The instrument is performing a
fractionation.
The instrument is currently not ready.
Possible reasons:
• the set temperature of the cooling
platform has not been reached
• top cover is open
The instrument has detected an error.
The instrument is not performing a
fractionation. The set temperature of
the cooling platform has been reached.
Comment
A running fractionation will be paused
if the top cover is opened.
A running fractionation will be aborted.
see
Table
on page 94
OFFGEL User Manual