Replace The Power Supply Assembly - GE Medical Systems Solar 9500 Service Manual

Information monitor
Table of Contents

Advertisement

Field Replaceable Units, Parts Lists and Drawings: Disassembly/Assembly
Assemble Front and Rear Covers

Replace the Power Supply Assembly

9-4
1. Lay the Solar 9500 processing unit on its front.
2. Place the rear housing cover and the mounting bracket on the back of
the unit being careful to align the screw bosses with the openings in
the plastic cover and install the four screws.
3. Turn the unit over.
4. Insert the top of the front cover into the hinges on the top of the rear
cover.
5. Rotate the front cover down until the two tabs engage and lock into
the rear cover.
&$87,216
The input capacitors store high voltages. After the supply
is turned off, bleeder resistor R57 discharges the
capacitors. It takes about 30 seconds for the voltage to
discharge to a safe level.
This assembly is extremely static sensitive and should be
handled using electrostatic discharge precautions.
1. Remove the front cover.
2. Loosen the two chassis mount screws and swing the power supply
chassis down to access interconnecting cables.
TO ACCESS POWER SUPPLY
Solar 9500 Information Monitor
2001085-061
LOOSEN 2 SCREWS
AND PULL DOWN
CONNECTORS
079A
Revision A