10% CO2 Calibration

Calibration Completion

Revision A
Module Calibration: Calibrate Mainstream CO2
7. Select Press when gas is supplied and the messages change to
Waiting for 10% Gas and Press when 10% CO2 is supplied...
8. Reconnect the calibration adapter to the tubing and apply 10%
calibration gas to the airway adapter of the sensor. When CAL gas is
applied during calibration, a waveform similar to that displayed
below shows the CO2 level.
127(
To view the CO2 waveform, close the Service Menu window and
move the CO2 calibration window.
9. Next, select Press when gas is supplied to begin 10% calibration.
A blanking out of both Module Status windows indicates that the
calibration was completed.
If the calibration failed one of the following messages appear:
One Of The Cal Gases is Wrong
or
Bad 10% Gas Used
127(
The previously stored calibration factors are held if the calibration
fails.
Solar 9500 Information Monitor
2001085-061
066A
5-11