Calibrate Sidestream Co2; About The Procedure; Manufacturer Recommendation; Test Equipment - GE Medical Systems Solar 9500 Service Manual

Information monitor
Table of Contents

Advertisement

Calibrate Sidestream CO2

About the Procedure

Manufacturer Recommendation

Test Equipment

Calibration Procedure

5-12
Module Calibration: Calibrate Sidestream CO2
This procedure provides a method of verifying the accuracy of the MGA-
IR Sidestream CO2 module.
The manufacturer recommends performing this procedure upon initially
receiving the module, before it is used on a patient, and once each year
thereafter. Also, perform the procedure each time the module is opened
for service or repair.
The following items are required to successfully complete the CO2
calibration procedure:
Description
CO2 Module Calibration Kit
The calibration procedure is relatively simple and should be performed
once per year. The calibration procedure is as follows:
1. Connect the Calibration Kit to the MGA-IR Sidestream CO2 Module
and Sensor as shown below.
2. Enter the Service Menu as described on page 5-3.
Solar 9500 Information Monitor
2001085-061
Part Number
405910-001
Quantity
1
067A
Revision A

Advertisement

Table of Contents
loading

Table of Contents