Calibrate Barometric Pressure for Interfaced CO2

About the Procedure

Procedure

Revision A
Module Calibration: Calibrate Barometric Pressure for Interfaced CO2
This procedure sets the barometric pressure for an interfaced CO2 using
Calibrate... in the Module Calibration window of the Service Menu.
127(
The buttons to display the calibration means are active only if the
associated Parameter is active and an interface device is present.
1. Enter the Service Menu as described on page 5-3.
2. Select Calibrate... to display the following window.
3. CO2... will be active when an interface device with CO2 parameter is
connected to the Solar 9500. Select CO2...
4. Enter the current pressure by using the up and down arrows, the
slide bar or touching the keypad icon and entering the pressure from
the keypad pop-up.
127(
Barometric pressure is used to calculate the percentage of CO2 in the
patient's airway from the mmHg value. This calculation is based on
the following equation: %CO2 = mmHg CO2 x 100% / mmHg
barometric pressure.
Solar 9500 Information Monitor
2001085-061
043A
108A
5-15