GE Medical Systems Solar 9500 Service Manual page 62

Information monitor
Table of Contents

Advertisement

4-10
Software Configuration: Monitor Service Menu
127(
After modifying any defaults remember to return to the Setup
Parameter Defaults window and select one of the following options.
Save – After modifying a parameter default settings, select Save to save
the changes.
Restore – Select Restore to restore the most recently saved settings.
When Restore is selected, a message window opens:
This will change all parameter settings back to the user default
settings. Do you wish to continue?
Select Yes to change the settings, or No to close the message window
without making changes.
Reset – Select Reset to reset all settings back to the original factory
settings. When Reset is selected, a message window opens:
This will change all parameter settings back to the original
GE clinical layouts default settings. Do you wish to continue?
Select Yes to change the settings, or No to close the message window
without making changes.
Print – Select Print to initiate a printout of the saved user default
settings for all of the available parameters.
GE Medical Systems Information Technologies recommends printing
default settings after making changes.
Close – Select Close to close the Setup Parameter Defaults window.
Solar 9500 Information Monitor
2001085-061
037A
Revision A