2.8
X-ray Generator (High Voltage Power Supply(ies))
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-66 System Overview
The PRODIGY X-ray production system combines a constant potential x-ray
generator (consisting of either one or two high voltage power supplies) and an
X-ray tube. The x-rya generator provides a continuous output of 38 kV each
and up to 5 mA (PRODIGY I) or 3.5 mA (PRODIGY II) to the X-ray tube. The
x-ray insert anode is composed of a tungsten alloy. The specifications of the
X-ray generation subsystem is outlined in section 2.16.7.
• The High Voltage Power Supply(ies) provide the anode/cathode potential
to the X-ray insert.
• PRODIGY I scanners use two high-voltage power supplies (±40kVDC).
During normal operation, 76kVp is applied (+38kV at the anode and -
38kV at the cathode).
• PRODIGY II scanners use a single high-voltage power supply
(±40kVDC). During normal operation, 76kVp is applied (+38kV at the
anode and -38kV at the cathode).
• The High Voltage Power Supplies are controlled by the (c)SBC.
• These power supplies are powered by the AC line voltage and have their
own built in fusing.
• The AC power is routed thru the isolation transformer (PRODIGY I Only)
and the supply(ies) are enabled via a relay (see section 2.1.1).
PRODIGY Service Manual (Rev C - 2000)
Need help?
Do you have a question about the PRODIGY and is the answer not in the manual?
Questions and answers