Field Replaceable Units, Parts Lists and Drawings: Disassembly/Assembly

Replace the NLX Motherboard

9-6
&$87,21
This assembly is extremely static sensitive and should be
handled using electrostatic discharge precautions.
1. Remove front and rear covers.
2. Loosen the two screws on power supply and swing the power supply
down as shown.
3. Loosen the five screws on the top plate and remove.
4. Lift up on the PCB lock to release and remove the NLX Motherboard
assembly.
5. Install the new NLX Motherboard assembly. Ensure that the board
EMI fingers are inside the chassis channel.
6. Check that the new motherboard is properly seated then move the
PCB lock to the lock position.
7. Replace the top plate and power supply then tighten all screws.
8. Replace the front and rear covers.
9. Reconnect the power cord and all communication cables. Test for
proper operation and electrical safety (see chapter 6, Maintenance).
Solar 9500 Information Monitor
2001085-061
TOP PLATE SCREWS
LOCK
( 5 PLACES)
RELEASE
ENSURE THAT BOARD
EMI FINGERS ARE
INSIDE CHANNEL.
POWER SUPPLY SCREWS
( 2 PLACES)
097A