Unit Defaults

Parameters...
Revision A
Software Configuration: Monitor Service Menu
Selecting the Parameters... option under Unit Defaults displays the Setup
Parameter Defaults window. All of the parameter menu options as well
as arrhythmia and parameter alarm limits, alarm levels and alarm
volume can be accessed, modified, and saved as unit defaults.
Setup Parameter Defaults
In the Setup Parameter Defaults window, there are buttons for each
available parameter. Select the appropriate button to open the
corresponding defaults window and view or modify the current default
settings for an available parameter. Refer to "Individual Parameter
Windows" on the following pages.
&$87,21
Incorrect Parameter Settings. Changing the
parameter defaults does NOT affect the current
parameter and alarm settings until you select Save, and
then Restore on the Setup Parameter Defaults screen.
127(
The difference between run-time changes and parameter default
changes is whether the parameter is accessed through the main
menu or the service menu. Parameter menus accessed from the main
menu, display "Control" in the upper-right corner whereas they
display "Defaults" via the service menu. For example, "ECG Control"
vs. "ECG Defaults."
Solar 9500 Information Monitor
2001085-061
034B
4-7