GE Medical Systems Solar 9500 Service Manual page 61

Information monitor
Table of Contents

Advertisement

A
A
Alarm Limits
B
Alarm Level and State
C
Alarm Volume
D
Defaults
Alarm Graphing
E
Arrhythmia Alarm Levels... Select Arrhythmia Alarm Levels to open the Arrhythmia
F
System Alarm Levels...
G
Revision A
Software Configuration: Monitor Service Menu
Set alarm limits on the left side of the screen by
selecting the up/down arrows.
Set alarm levels to Crisis, Warning, Advisory or
Message. Set alarm state to On, Off, or Always On.
NOTE: Setting an alarm state to Always On means
Set alarm volume by selecting the up/down arrows. Set
alarm volume to On, Off, or Always On.
NOTE: Setting the alarm volume to Always On means
Select Restore... to restore alarm defaults to either User
Defaults or GE clinical layouts Defaults.
Set alarm graphing to On or Off to enable or disable
automatic graphing of alarms as they occur.
Defaults window. All 15 monitored arrhythmia
conditions have alarm level choices Crisis, Warning,
Advisory or Message.
Select System Alarm Levels... to open the System Alarm
Defaults window. Set ECG Leads Fail and SpO2 Probe
Off Patient to System Warning or System Advisory.
Solar 9500 Information Monitor
2001085-061
B
that the user does not have the option to change
the individual parameter alarm to the Off state.
the user does not have the option to turn the
alarm volume off.
C
D
E
F
G
036A
4-9