Field Replaceable Units, Parts Lists and Drawings: Disassembly/Assembly
9-16
Only one software configuration type can be activated at a time. After
selecting, you will be queried if you really want to make the selection as
it restarts the system.
&$87,21
Bed layouts, Waveform layouts, and Parameter defaults
can only be activated against the same software revision
that was running on the machine that exported the
layouts.
Bed layouts
When bed layouts are activated:
a pop-up menu displays whether the activation was successful or not,
the system is restarted,
the new bed layouts appear in the switch layout menu,
if the current bed layout display was updated, the display changes
when the system is restarted, and
old layouts with the same name are overwritten and cannot be
recovered.
Waveform layouts
When waveform layouts are activated:
a pop-up menu displays whether the activation was successful or not,
the system is restarted,
the new waveform layouts appear in the window layout list of the
"Configure Window" menu, and
old layouts with the same name are overwritten and cannot be
recovered.
Parameter defaults
When parameter defaults are activated:
a pop-up menu displays whether the activation was successful or not,
the system is restarted,
the system switches to the newly activated parameter settings, and
the parameter setup becomes the unit default.
127(
The original settings are overwritten and cannot be recovered.
Solar 9500 Information Monitor