Field Replaceable Units, Parts Lists and Drawings: Disassembly/Assembly
9-14
3. Select a unit name and then the beds in the unit to receive the
network service. Or enter the Internet Address (e.g. 126.1.2.3) to
identify the receiving sites.
4. Select the configuration type (Bedlayouts, Waveform layouts,
Parameter defaults, or Software) to be exported. Only one type of
configuration type can be selected for export at a time.
If Bedlayouts is selected, a list of all user defined bed layouts
currently present on the exporting system displays.
If Waveform layouts is selected, a list of all waveform layouts
currently present on the exporting system displays.
If Parameter defaults is selected, the Unit default for all
parameters as well as alarms is sent.
If Software is selected, a list of all the currently installed
software releases on the exporting system displays.
127(
The software consists of three types: S9500 OS, System Apps., and
Monitor Apps. For proper operation all three software types must be
present at each Solar 9500. Each software release clarifies which
versions of all three software types are required. The software is
checked for required versions before being activated.
5. Select Export to queue the sending of all configurations that have
been selected from the list or Cancel to close the window.
6. After an export is completed, a window listing the success/failure
status of each export displays. You can now choose to Print or
Dismiss (close) the window.
Solar 9500 Information Monitor
2001085-061
057A
Revision A