Processing Unit Setup

Revision A
Installation: Processing Unit Setup
Check the unit for proper voltage setting before installation. Although
the AC mains voltage on Solar 9500 processing unit is factory set for your
requirements, the power supply could be damaged if S1 is in the wrong
position.
&$87,21
Make sure the voltage selection switch is in the correct
position before connecting AC mains. The power supply
will be damaged if the switch is in the 115 volt position
and a high range voltage (195-270VAC) is applied. The
supply will not function if in the 230 volt position when
the low range voltage (90-135VAC) is applied.
1. Lay the Solar 9500 processing unit on its back.
2. Simultaneously press in on the two release tabs and lift up the front
cover.
3. View the power supply voltage setting through the window at the top
of the unit.
Solar 9500 Information Monitor
2001085-061
007A
008A
3-3