When automatically starting the RF output in
lesion mode with the impedance of probe 2 open
circuit.
When automatically starting the RF output in
lesion mode with the impedance of probe 3 open
circuit.
When automatically starting the RF output in pulse
mode with the impedance of probe 1 open circuit.
When automatically starting the RF output in pulse
mode with the impedance of probe 2 open circuit.
When automatically starting the RF output in pulse
mode with the impedance of probe 3 open circuit.
When manually starting the RF output in any
mode with the impedance of probe 1 less than 50
Ohms
When manually starting the RF output in any
mode with the impedance of probe 2 less than 50
Ohms
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
®
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press lesion to go back to procedure screen
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press lesion to go back to procedure screen
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
Unable to deliver RF power
Impedance open circuit on probe 2.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
Unable to deliver RF power
Impedance open circuit on probe 3.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
Unable to deliver RF power
Impedance is lower than 50 Ohms on probe 1.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
Unable to deliver RF power
Impedance is lower than 50 Ohms on probe 2.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
10September2012 Issue 06