When carrying out treatment take care to avoid the following risks: -
RISK OF RF BURNS TO PATIENT - Ensure patient does not come into contact with metal parts of the
table and its accessories – antistatic sheeting recommended.
RISK OF RF BURNS TO PATIENT - Avoid skin to skin contact between different parts of patient's body
(for example between the arms and the body of the patient) – use dry gauze if necessary.
RISK OF RF BURNS TO PATIENT: - Avoid using physiological monitoring equipment during a procedure
– if monitoring is required, monitoring electrodes should be placed as far as possible from the
NeuroTherm cannula. Monitoring devices which use needle electrodes are not recommended.
RISK OF RF BURNS TO PATIENT: - Position all cables to the NeuroTherm cannula and dispersive plate
in such a way to avoid contact with the patient or other leads.
INTERFERENCE WITH ACTIVE INPLANTS: - Check whether the patient has a cardiac pacemaker or
other active implant. A possible hazard exists because interference with the action of the pacemaker may
occur or the pacemaker may be damaged. In case of doubt, obtain qualified advice.
INTERFERENCE WITH OTHER EQUIPMENT: - During RF Lesioning procedures the radiated electrical
fields may interfere with other electrical medical equipment. (See Section 3.4 to Minimise
Electromagnetic Interference)
GENERAL CONSIDERATIONS: - Regularly inspect the accessories of the NeuroTherm, in particular
electrode cables should be checked for possible damage to the insulation.
DO NOT USE ENDOSCOPE: - The accessories are not appropriate for endoscopic use.
USE OF FLUIDS: - Ensure that if fluids (saline etc.) are being used during a procedure they should be
positioned away from the NeuroTherm.
RISK OF RF BURNS TO PATIENT: - In Manual Lesion Mode select the lowest possible power for
intended purpose.
RISK OF RF BURNS TO PATIENT: - Check the Dispersive (Neutral) Lead and the Dispersive Pad before
applying power to the patient.
PROBES: - Use only NeuroTherm
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
TM
, Radionics Disc Trode
®
TM
TM
, Spinecath
, or ACUTHERM
10September2012 Issue 06
2-2
TM
probes.