In the centre of the screen there is a graph similar to the other RF modes, showing temperature
plotted against time. At the top left of the screen a counter shows the number of pulses left in the
procedure. This will count down until all of the pulses have been supplied to the patient, the alarm
will sound. The procedure is now complete. Press "end session" to return to the "Procedure
Setting" screen.
Note In pulse Dose mode The NT1100 will deliver full 20ms Pulses of 45 volts, should the
temperature exceed 42ºC the pulse will not be delivered and the counter will stop until the
temperature is below 42ºC. The count will resume at this point, to ensure the correct amount of
energy is delivered.

Two electrodes RF

1. Two Electrode RF is only available in Auto mode. It is selected at the "Procedure Setting"
screen menu. Highlight the choice and press 'Accept'. (Each electrode uses the Dispersive lead
as a return path for the RF signal).
2. After the procedure selection screen Press "Accept" the NT1100 screen will automatically
display sensory and motor Stimulation. Each electrode has its own graphical area.
Electrode 1 on the left, with electrode 2 on the right. Use the
between electrode 1 and electrode 2. The log functionality is the same as that of a single
electrode mode.